Autoclaving of Bottle Top Dispenser

Autoclaving of Bottle Top Dispenser

Starlab Scientific Co., Ltd’s bottle top dispenser withstands steam sterilization at 121 °C, 2 bar absolute (15 psi) for 20 minutes in accordance with widely used guidelines. It is the user’s responsibility to ensure effective autoclaving.
Preparation prior to the autoclaving of the dispenser

Autoclaving of Bottle Top Dispenser1             Autoclaving of Bottle Top Dispenser2

Step 1: Please make sure leaning the dispenser before sterilization operation
Step 2: Please remove the discharge tube with removed closure cap, filling tube, and put all parts on a towel. Please note here must avoid contact with hot metal surfaces
Step 3: Please lay the instrument with the piston in down position on the towel
Step 4: All parts can be autoclaved now

Please note:
Must make sure allow to cool down in the closed autoclave to avoid deformation through too rapid cooling
After every autoclaving, all parts should be inspected in case that deformities or damage. replace them when necessary
Do not reassemble the components of bottle top dispenser until it has cooled down to room temperature (The cooling time shall be approx. of 2 hours)
All the mounting tool dispenser can be steam sterilized at 121 °C

Availability:
Mechanical Bottle-Top Dispenser
Range: 0.5-5mL, 1-10mL, 2.5-25mL, 5-50mL
Electronic Bottle-Top Dispenser
Range: 0.1-99.9mL